FLUIGENT-logo

FLUIGENT ODROPSEQFPCK Bidirectional Microfluidic Flow Sensor

FLUIGENT-ODROPSEQFPCK-Bidirectional-Microfluidic-Flow-Sensor-fig-1

Fa'amatalaga o oloa

  • Igoa o oloa: DropSeq Fluidic
  • Numera Vaega: ODROPSEQFPCK
  • Fa'aliliuga: Oketopa 20

Fa'amatalaga o oloa

The DropSeq Fluidic protocol aims to explain how to replicate the fluidic part of the Macosco DropSeq protocol using Fluigent pressure controllers instead of syringe pumps. The advantage of pressure controllers is their fast response times, enabling the generation of droplets quickly. Pressure control allows for droplets of the right size to be produced in less than two seconds. Additionally, flow sensors enable direct flow control, mimicking the operation of a syringe pump.

Materials and Methods:
To implement this protocol, you will need the following materials:

  • DropSeq package
  • External product (optional)
  • Various components as listed in the content description

Fa'atonuga o le Fa'aaogaina o Mea

Preparing the Device for a New Experiment:

  1. Check the PDMS devices before use.
  2. Peel off a small area of the film covering the generators.
  3. Add a drop of dSURFan  or alternative surfactant into the outlet and wait.
  4. Connect the tubing to the microfluidic device.
  5. Apply a small pressure using Flow-EZ.
  6. Wait for fluid to come out of the in-/outlets and set pressure back to 0 mbar.
  7. Start vortex stirrers and apply pressure to reservoirs until fluid comes out.
  8. Connect all tubings according to the schematic.

Launching an Experiment:
After proper connection, start a DropSeq run by applying pressures to different reservoirs using A-i-O software. Droplets should form in a few seconds.

Flowrate Control:
With a Flow Unit, use the Flow-EZ as a syringe pump. The A-i-O software allowflow ratete commands, adjusting pressures accordingly. Save the pressure configuration once set.

SINI O LE TULAFONO

  • This protocol aims to explain how to reproduce the fluidic part of the Macosco DropSeq protocol using Fluigent pressure controllers instead of syringe pumps.
  • O le advan aututage of pressure controllers is the very fast response times. Using pressure control, droplets of the right size can be generated in less than two seconds. Using flow sensors, it is also possible to use direct flow control in order to use the pressure controller in the same way you would use a syringe pump.

METOTIA MA METOTIA

To realize this protocol, you need the following material:

Product                    Reference              Content description
DropSeq package ODROPSEQFPCK 3 *FlowEZ 2 pa
1 * Sootaga
1 * Pepa 15 ml
2 *Pcap 2ml
2 * Fa'asolo vaega M
1 * Fa'asuavai iunite L
DropSeq PDMS chip containing 23 designs
Flow EZ sapalai pusa
2*Low flow rate kit
Pusa maualuga tafe
DropSeq connecting and tubing kit
Oloa e filifili ai IMCA001 Digital high-speed microscope
DR-RE-SU-12 12mL dSURF
FLPG+ Punavai omiga
External product Barcoded beads Barcoded beads: Chemgenes CSO-2011*
Vortex fa'afefiloi Vortex mixer: Sigma, Fisher Scientific…

The first step is to recreate the following setup.

FLUIGENT-ODROPSEQFPCK-Bidirectional-Microfluidic-Flow-Sensor-fig-2

FUAFUAGA

Preparing the device for a new experiment
When starting a new experiment, it is recommended to push out all the air from the chip and the tubing. For that, follow the following steps:

  1. Before using the PDMS devices, please check the generators carefully. If you have any concerns/questions, please do not hesitate to contact us.
  2. Do not fully peel off the film at the beginning: Please cut off the small area for each generator for use. (All of our PDMS chips are covered with Sealing Film, hydrophobic porous sealing film with medical-grade adhesive for tissue culture plates.)
  3. Add one drop of dSURF (or Fluorinated oil with other surfactant if not using the dSURF) into the outlet before starting the experiment and wait for a few minutes. Please increase the flow rates of the F-oil and aqueous phases step by step. Avoid jumping from 0 to ~14 mL/h (or maximum flowrates) directly.
  4. Connect the dSURF (or Fluorinated oil with other surfactant if not using the dSURF) tubing to the right entry port on the microfluidic device.
  5. Apply a small pressure using the Flow-EZ (about 200 mbar should be enough)
  6. Wait until you see some oil coming out of the open in-/outlets of the chip and then set the pressure back to 0 mbar.
  7. Start the vortex stirrers and apply a small pressure to the two other reservoirs until you see a small drop of fluid coming out of the end of the tube.
  8. Stop all pressures and connect all tubings according to the schematic

Launching an experiment
Once the device is connected properly, starting a DropSeq run is very fast. Simply apply the following pressures to the different reservoirs using the A-i-O software. The droplets should appear in less than a few seconds.

  • P (Oil) = 1300 mbar
  • P (Cells) = 540 mbar
  • P (Beads) = 580 mbar
    These pressures should generate droplets of about 90-130 µm, which leads to encapsulation rates from 1 to 3 ).
    Motugaafa:
    Using the A-i-O software, pressure configurations can easily be saved. This allows setting all pressures to a given point at the same time.

Flowrate control

  • Combined with a Flow Unit, the Flow-EZ can be used exactly as a syringe pump. The A-i-O software allows flowrate commands. A feedback loop on the flowrate then adjusts the pressures to the right value. This technique can be useful to evaluate the necessary pressures for the experiment.
  • Once in the right state, the pressure configuration can then be saved.

    FLUIGENT-ODROPSEQFPCK-Bidirectional-Microfluidic-Flow-Sensor-fig-3

For additional information, contact us by email: contact@fluigent.com or consult our webnofoaga: www.fluigent.com.

FAQ

How fast can droplets be generated using pressure controllers?

Droplets of the right size can be generated in less than two seconds using pressure controllers.

Can I use direct flow control with this protocol?

Yes, flow sensors enable direct flow control similar to using a syringe pump.

Pepa / Punaoa

FLUIGENT ODROPSEQFPCK Bidirectional Microfluidic Flow Sensor [pdf] Tusi Taiala
ODROPSEQFPCK, IMCA001, DR-RE-SU-12, ODROPSEQFPCK Bidirectional Microfluidic Flow Sensor, ODROPSEQFPCK, Bidirectional Microfluidic Flow Sensor, Microfluidic Flow Sensor, Flow Sensor

Fa'asinomaga

Tuu se faamatalaga

E le fa'asalalauina lau tuatusi imeli. Fa'ailogaina fanua mana'omia *