Tektronix AFG1022

Tusitaiala mo le Fa'aaogāina o le Tektronix AFG1022 Arbitrary Function Generator

Model: AFG1022

1. Folasaga

The Tektronix AFG1022 Arbitrary Function Generator is a versatile instrument designed for generating various waveforms required in laboratory and testing environments. With a bandwidth of 25 MHz or 60 MHz and dual output channels, it provides an output amplitude ranging from 1 mVpp to 10 Vpp across its full bandwidth. This device offers a comprehensive set of features for waveform generation, making it a valuable tool for a wide range of applications.

O mea taua e aofia ai:

  • Dual-channel operation with 25 MHz bandwidth.
  • Tuuina atu amplitude from 1 mVpp to 10 Vpp.
  • 14-bit vertical resolution and 1 µHz frequency resolution.
  • 50 built-in arbitrary waveforms.
  • Support for continuous, modulation, sweep, and burst modes.
  • Dedicated 3.95-inch TFT LCD display for clear visualization.
  • User-friendly front panel and menu navigation.
Luma view of the Tektronix AFG1022 Arbitrary Function Generator with its display showing waveform settings.
Ata 1.1: Luma view of the Tektronix AFG1022 Arbitrary Function Generator. This image displays the front panel of the AFG1022, highlighting the large color display, control knobs, function buttons, and output connectors. The display shows settings for sweep time, amplitude, and offset, along with various waveform types like sine, square, ramp, pulse, arbitrary, and noise.

2. Fa'amatalaga Saogalemu

Before operating the Tektronix AFG1022, please read and understand all safety instructions to prevent injury and avoid damage to the instrument or other connected equipment. This device is designed for use by qualified personnel only.

Puipuiga Lautele Saogalemu:

  • Fa'avae: Always ensure the instrument is properly grounded. The power cord provided has a three-prong plug for this purpose. Do not defeat the grounding-type plug.
  • Punavai Malosi: Connect the instrument to a power source that is within the specified voltage range (100-240V AC, 50/60 Hz). Verify the line selector switch on the rear panel is set correctly for your local voltage.
  • Savili: Ensure proper ventilation around the instrument. Do not block ventilation openings. Overheating can cause damage.
  • Siosiomaga: Operate the instrument in a clean, dry environment. Avoid exposure to moisture, extreme temperatures, or corrosive gases.
  • Au'aunaga: Refer all servicing to qualified service personnel. Do not attempt to service the instrument yourself unless you are qualified to do so.
  • Uaea ma feso'ota'iga: Use only specified cables and connectors. Ensure all connections are secure before applying power.
  • Fa'amamaina: Motusi le eletise a'o le'i fa'amamāina. Faaaoga se ie vaivai ma mago. Aua le fa'aaogaina ni mea fa'amamā vai po'o ni mea fa'amamā ea.

3. O a mea oi totonu o le Pusa

Upon unpacking your Tektronix AFG1022 Arbitrary Function Generator, please verify that all the following items are included:

  • Tektronix AFG1022 Arbitrary Function Generator Unit
  • Fa'amatalaga o le Tausi
  • CD (containing software and documentation)
  • Tusi Lesona a le Tagata (lenei pepa)
  • Fuse (2 pieces, spare)
  • Cables (3 pieces, including USB and BNC cables)
  • Fa'ailoga Fa'ailoga
  • Uaea Malosi

If any items are missing or damaged, please contact your Tektronix supplier or customer support immediately.

4. Seti

Follow these steps to set up your AFG1022 Arbitrary Function Generator for initial use.

4.1 Tatalaina ma Asiasiga

  1. Carefully remove the instrument and all accessories from the packaging.
  2. Inspect the instrument for any signs of physical damage that may have occurred during transit. If damage is found, contact your supplier.
  3. Retain the original packaging materials for future transport or storage.

4.2 So'oga Malosi

  1. Locate the power inlet on the rear panel of the AFG1022.
  2. Faamaonia le Filifili Laina switch on the rear panel is set to the correct voltage for your region (110V AC or 220V AC). Adjust if necessary.
  3. Connect the provided power cord to the instrument's power inlet and then to a grounded AC power outlet.
Tua view of the Tektronix AFG1022 Arbitrary Function Generator showing input/output ports and power connections.
Ata 4.1: Tua view of the Tektronix AFG1022. The rear panel shows the Ref Clk Out, Ext Trig/Burst/Fsk In, Ext Mod In, and USB ports. It also includes the AC power inlet with fuse, and a line selector switch for 110V/220V AC operation. A warning about proper grounding is visible.

4.3 Connecting Outputs

  1. Identify the two output channels (Out1 and Out2) on the front panel. These are BNC connectors.
  2. Connect appropriate BNC cables from the desired output channel(s) to your test circuit or device under test.
  3. Ensure the impedance setting on the AFG1022 matches the impedance of your load (typically 50Ω or high impedance).

5. Fa'atonuga o Galuega

O lenei vaega o loʻo tuʻuina atu se faʻaopoopogaview of the basic operation of the Tektronix AFG1022. For detailed instructions on specific functions, refer to the comprehensive user manual provided on the included CD.

5.1 Powering On and Initial Screen

  1. Press the power button located on the front panel to turn on the instrument.
  2. The display will show the Tektronix logo and then transition to the main waveform generation interface.

5.2 Basic Waveform Generation

The AFG1022 can generate various standard waveforms and arbitrary waveforms.

  • Selecting Waveform Type: Use the dedicated buttons below the display (Sine, Square, Ramp, Pulse, Arb, Noise) to select the desired waveform.
  • Setting Frequency: Use the numeric keypad and the rotary knob to input the desired frequency. The display will update in real-time.
  • Faatulagaina Amplitude and Offset: Fetuunai le amplitude (Vpp) and DC offset (V) using the corresponding soft keys and the rotary knob.
  • Fa'agaoioi Galuega: Oomi le Ki/Peti button for Channel 1 (Out1) or Channel 2 (Out2) to enable the waveform output. The button will illuminate when the output is active.
Fa'afa'afa'o i luma view of the Tektronix AFG1022 Arbitrary Function Generator on a workbench.
Ata 5.1: Fa'afa'afa'o i luma view of the Tektronix AFG1022. This image provides an angled view of the AFG1022 on a workbench, emphasizing its compact design and the clarity of its display. The display shows a sweep time setting of 25 billion seconds, indicating its wide range of capabilities. Various tools and other lab equipment are visible in the background.

5.3 Galuega Maualuga

  • Fa'atonu: Oomi le Mod button to access modulation settings (AM, FM, PM, FSK, PWM).
  • Salu: Configure frequency sweeps by pressing the Salu ki malu
  • Pa'u: Generate burst waveforms by pressing the Pa ki malu
  • Fa'asa'o Fa'agalu: The AFG1022 supports user-defined arbitrary waveforms. Refer to the full manual for instructions on creating and loading custom waveforms.

6. Tausiga

Proper maintenance ensures the longevity and optimal performance of your Tektronix AFG1022.

6.1 Fa'amamāina

  • Always disconnect the power cord before cleaning the instrument.
  • Use a soft, dry cloth to wipe the exterior surfaces. For stubborn dirt, a cloth lightly dampened with mild detergent and water may be used.
  • Do not use abrasive cleaners, solvents, or strong chemicals, as these can damage the plastic parts and labels.
  • Avoid spraying liquids directly onto the instrument.

6.2 Suiga Fuse

If the instrument does not power on, check the fuse located in the power inlet on the rear panel. Ensure the replacement fuse matches the specified type and rating (e.g., 250V, 2A). Refer to the full manual for detailed instructions on fuse replacement.

6.3 Fa'avasegaina

For continued accuracy, periodic calibration by qualified service personnel is recommended. Refer to the calibration certificate for the recommended calibration interval.

7. Faʻafitauli

This section provides solutions to common issues you might encounter with your AFG1022. If the problem persists, contact Tektronix customer support.

Fa'afitauliMafuaaga e ono tupuFofo
E le ola le meafaigaluega.No power, faulty power cord, blown fuse.Check power cord connection. Verify power outlet. Check and replace fuse if necessary. Ensure line selector is correct.
No waveform output.Output channel not enabled, incorrect amplitude/offset settings, faulty cable, incorrect load impedance.Press the On/Off button for the desired channel. Verify amplitude and offset settings. Check BNC cable connections. Ensure load impedance matches instrument setting.
E avanoa pe fa'aaisa le fa'aaliga.Software error, internal fault.Power cycle the instrument (turn off, wait 10 seconds, turn on). If issue persists, contact support.
Incorrect waveform shape/frequency.Incorrect settings, external interference.Double-check all waveform parameters (frequency, amplitude, offset, modulation settings). Ensure proper shielding of cables.

8. Fa'amatalaga

The following table lists the key specifications for the Tektronix AFG1022 Arbitrary Function Generator. All specifications are typical unless otherwise noted.

ParameterTaua
Igoa Fa'ata'ita'igaAFG1022
Bandwidth25 MHz
AualaAuala lua
Tuuina atu Amplotoa1 mVpp i le 10 Vpp
Iugafono Vertical14-bit
I'uga o taimi1 µHz
Built-in Waveforms50 Arbitrary Waveforms
Fa'aaliga3.95-inisi TFT LCD
Punavai MalosiElectricity Powered
Voltage100-240V AC (selectable)
Mea mamafa4.9 Kilokalama (10.8 Pauna)
Fuafuaga Mea17.2 x 14 x 11.4 inisi
TufugaTektronix
Aso Muamua MauaIulai 15, 2015

9. Tuutuuga & Lagolago

Tektronix products are designed for reliability and performance. For specific warranty terms and conditions applicable to your AFG1022, please refer to the warranty information provided with your purchase or visit the official Tektronix webnofoaga.

Lagolago tagata fa'atau:

If you require technical assistance, have questions about your instrument, or need to arrange for service, please contact Tektronix customer support. Support contact information can typically be found on the Tektronix web'upega tafaʻilagi poʻo totonu o faʻamaumauga e aofia ai ma lau oloa.

When contacting support, please have your instrument's model number (AFG1022) and serial number ready. The serial number is located on the rear panel of the instrument.

For the latest information, drivers, software, and additional resources, please visit the official Tektronix webnofoaga: www.tek.com

Pepa Fa'atatau - AFG1022

Preview Tektronix AFG3000 Series: 25 Applications of Arbitrary/Function Generators
Explore 25 practical applications for Tektronix AFG3000 Series Arbitrary/Function Generators, covering embedded systems, RF, automotive, medical, industrial, and research fields. Enhance your electronic testing and design workflows.
Preview Tektronix KickStart Software Datasheet: Accelerate Measurement Efficiency
Explore the Tektronix KickStart software, a powerful PC application designed for quick instrument setup, data visualization, and analysis. Discover its various apps for DMMs, AFGs, data logging, I-V characterization, power supplies, oscilloscopes, battery simulation, and high resistivity measurements. Streamline your testing workflow and gain faster insights into your data.
Preview AWG70001A および AWG70002A 任意波形ジェネレータ インストールおよび安全に関する取扱説明書
Tektronix AWG70001A および AWG70002A 任意波形ジェネレータの設置と安全に関する取扱説明書。セットアップ、安全上の注意、および基本的な操作ガイダンスを網羅しています。
Preview AWG70001A および AWG70002A 任意波形ジェネレータ インストールおよび安全に関する取扱説明書
Tektronix AWG70001A および AWG70002A 任意波形ジェネレータの安全な設置、操作、および保守に関する包括的なガイド。安全上の注意、セットアップ手順、およびデバイス機能の概要を提供します。
Preview Tektronix PPG4001 PatternPro Programmable Pattern Generator User Manual
User manual for the Tektronix PPG4001 PatternPro Programmable Pattern Generator, detailing features, specifications, operation, safety, and remote programming for advanced signal generation.
Preview Tektronix 1410 NTSC Generator Instruction Manual
Comprehensive instruction manual for the Tektronix 1410 NTSC Generator, detailing operation, specifications, installation, theory of operation, maintenance, and troubleshooting for this video test equipment.